Safety And Efficacy Of Ascending Aortic Cannulation In Minimally Invasive Mitral Valve Surgery
Hiroyuki Nakajima, Akitoshi Takazawa, Masato Tochii, Akhiro Yoshitake, Toshihisa Asakura, Atsushi Iguchi.
Saitama International Medical Center, Saitama, Japan.
OBJECTIVE:Femoral artery cannulation has been generally utilized in minimally invasive cardiac surgery (MICS). However, the risk for aortic dissection may increase, comparing with standard aortic cannulation. We sought to delineate efficacy and safety of aortic cannulation via right pleural cavity.
METHODS:We retrospectively reviewed clinical records of 52 MICS operation performed at our institutions between December 2012 and December 2017. Aortic cannulation has been performed for 16 patients in group AO, while arterial cannula was connected with 8mm artificial graft, with was anastomosed with the femoral artery for 36 patients in group FA. As surgical technique of aortic cannulation, pursestring suture was made on the ascending aorta, cannula was introduced into right pleural cavity at 4th intercostals space on the mid-axillary line through a separate 5mm skin incision. Cannula was inserted into the aorta with the use of the guide wire.
RESULTS:There was no mortality, nor aortic dissection. Blood pressure in the arterial cannula was 200+/-26mmHg in group AO, which was significantly lower than 231+/-36mmHg in group FA (p=0.01). The amount of the flow was comparable (p=0.24). The skin incision was 55mm at chest and about 20mm at groin in group FA, while 53mm at chest and additional 5mm for cannula in group AO. CONCLUSIONS:Aortic cannulation could be safely performed through acceptable length of skin incision, and provides lower pressure in the arterial cannula, and minimizes the risk of aortic dissection, as comparable as the standard cardiac operation.
Back to 2018 Display Posters