Propensity matched comparison of endoclamping versus direct aortic clamping for minimally invasive mitral valve surgery
Joerg Kempfert, Alexander Meyer, Ayse Cetinkaya, Mani Arsalan, Tibor Ziegelhoeffer, Markus Schoenburg, Thomas Walther.
Kerckhoff Clinic, Bad Nauheim, Germany.

OBJECTIVE: To compare endoaortic clamping (Intraclude™, Edwards Lifesciences, Irvine, CA, USA) versus conventional clamping (Chitwood clamp) during minimally invasive mitral procedures by means of propensity score matching.
METHODS: Amongst more than 600 consecutive minimally invasive mitral valve procedures the IntraClude3 (Edwards Lifesciences) was used in 43 patients. To adjust for differences in baseline variables propensity score (PS) based matching was performed. The PS was modelled using logistic regression with 41 preoperative covariates consisting of clinical and demographical data.
RESULTS: Propensity score matching resulted in 37 matched pairs (matching efficiency 86%) and resulted in a reduction of standardized mean difference almost of all baseline variables. There were no significant differences of baseline variables left after matching.
In these 74 patients a 100% reconstruction rate could be achieved. There were no deaths or strokes observed at 30-days after surgery. Endoclamp cannulation and insertion was performed without any vascular complication. Procedure time (Endoclamp 189±32 vs. 189±63min, p=0.82), CPB time (Endoclamp 134±27vs. 120±53min, p=0.35) and X-clamp time (Endoclamp 83±20min vs. 75±35min p=0.50) were similar amongst groups. ICU stay (Endoclamp 24±27h vs. 23±13h, p=0.12) and ventilation time (Endoclamp 9±4h vs. 8±4h, p=0.08) did not differ.
Conversion to sternotomy was required in one patient in each group. Rate of re-thoracotomies (Endoclamp 2 (5%) vs 1 (3%), p=1.0) and transfusions (p=0.33) did not differ between both techniques.
CONCLUSIONS: Minimally invasive mitral valve surgery is safe and reliable using both options for aortic occlusion and subsequent cardioplegic arrest. Despite the fact that our patient numbers were rather small, we did not observed any relevant differences in clinical and functional outcome. Overall outcomes of minimally invasive mitral valve surgery are good with low additional risks.