Rapid Deployment Aortic Valves Deliver Superior Hemodynamic Performance
Lisong Ai1, Harvey Chen1, Virginia Lin1, Vinayak N. Bapat2.
1Edwards Lifesciences LLC., Irvine, CA, USA, 2Guys and St. Thomas' Hospital NHS Trust, London, United Kingdom.

OBJECTIVE: Clinical studies have demonstrated superior hemodynamic performance with Rapid Deployment Aortic Valve Replacement (RDAVR), with accelerated left ventricular mass regression and enhanced functional status and survival. These valves employ an expandable frame to ensure proper valve seating and annular sealing. To help explain these clinical findings, in vitro flow system and Particle Image Velocimetry (PIV) flow visualization were used to analyze the contribution of frame expansion on valve hemodynamic performance.
METHODS: The control valve was a 19 mm conventional surgical valve and the test valves were 19 mm rapid-deployment valves. Dimensions for the elliptical annulus and LVOT were derived from clinical 3D multi-slice computer tomography (MSCT) reconstructions and were scaled to a perimeter of 19 mm. Elastomeric material was used to compose the simulated aortic valve annulus. All valves were positioned within the annulus using the commissure orientation from the MSCT images. Rapid-deployment valve frames were then expanded within the annulus. In accordance with the official valve testing standards for cardiac valve prostheses (ISO 5840) each valve was placed in the steady forward flow system using a blood analog solution for flow visualization and hemodynamic testing at six different flow rates.
RESULTS: Compared to the control surgical valve, frame expansion was associated with an increase in effective orifice area (6.3 ± 0.7%) and a decrease in mean valvular pressure gradient (-15.0 ± 1.6%). PIV demonstrated that the rapid-deployment valves were associated with a lower maximum velocity (-5.2 ± 1.2%), lower maximum shear stress (-14.5 ± 11.4%), and lower turbulent kinetic energy (-11.1 ± 9.7%).
CONCLUSIONS: This study confirmed the favorable clinical findings that valves designed for RDAVR deliver superior hemodynamic performance. Further, PIV demonstrated that this superior performance may be attributed to the ability of rapid-deployment valves to transform an elliptical LVOT into circular one, unlike conventional surgical valves, which are associated with an abrupt transition from the elliptical LVOT to circular valve flow area.