Late follow-up of a single center experience with a novel device for Apico-aortic conduit implantation
Gabriele Di Giammarco, Carlo Canosa, Massimiliano Foschi, Daniele Marinelli, Michele Di Mauro, Donato Micucci.
University "G.D'Annunzio", Chieti, Italy.
OBJECTIVE: High-risk patients affected by severe aortic stenosis represent an increasing population. The aortic valve bypass (AVB) in selected patients is a safe alternative to conventional aortic valve replacement. The introduction of an automated system for left ventricular coring makes the procedure more reproducible and easier. We report our experience focused on clinical follow-up.
METHODS: From March 2012 to August 2014, 20 patient were submitted to aortic-valve bypass(AVB)to treat a severe aortic valve stenosis using a composite conduit incorporating a bioprosthetic valve deployed by an automated system. Patients median age was 78 (25th-75th perc:71-81)years(40% males).Median ejection fraction(EF) was 55% (25th-75th perc:48-61).Median preoperative NYHA class was 3(25th-75th perc:2,8-3).The median LogEUROscore was 11,29 (25th-75th perc:7,8-20,4).Median STS morbidity/mortality risk was 23,24%(25th-75th perc:19,3-27,8).All the patients received AVB through a mid-lateral thoracotomy.In two patients a LIMA-to-LAD anastomosis was performed off-pump; a lung tumor resection was added in one of these. In another one atypical lung resection was performed.In two patients lung function was supported by ECMO during the deployment.In the last five cases we tied the sutures using an automated fastener.
RESULTS:No intraoperative deaths occurred.Median ICU stay and intubation time were 42(25-75°perc:23-59.5)and 17,5(25-75°perc:16-20,7)hours respectively.The median hospital stay duration was 13 (25th-75th perc:7,7-18) days.No cerebrovascular events occurred.There were 5 in-hospital deaths (20%): MOF (2),myocardial ischemia (1),acute gastrointestinal bleeding (1)and sepsis from Klebsiella pneumonia (1).Two patients died during follow-up; one for silent hepatocarcinoma(2months)and one for necrotic pancreatitis(6 months).Thirteen patients (median follow-up:640 days;91-820) are alive;12 patients are in NYHA class 1;one patient switched from NYHA 4 to 3.Actuarial survival and event-free survival is 77,5%.In the last five cases the use of automated fastener significantly reduced operation time from 217 to 155 minutes (p=0,026).
CONCLUSIONS: To the best of our knowledge this is the largest single center experience of AVB using this novel device. It may represent an alternative to conventional aortic valve replacement in high risk patients.The use of an automated system for knot tying improves operative time and standardization of the procedure with the chance to reduce the invasiveness of the operation.