A Novel Minimal Access Cardiac Surgery Automated Suturing Device With a Needle Sheath to Minimize the Risk of Needle-Stick Injuries
Devang J. Joshi1, Joshua K. Wong1, Jude S. Sauer2, Peter A. Knight1.
1University of Rochester Medical Center, Rochester, NY, USA, 2LSI Solutions, Victor, NY, USA.

OBJECTIVE: Despite advances in modern cardiac surgery, the threat of injury to surgeon and staff remains real. The complexity of cardiac surgery currently requires numerous needles. The risk of exposed “sharps,” especially contaminated surgical needles, is a known cause of infectious disease transmission. An automated, ergonomic and highly effective remote suturing device offering a protective sheath to shield suture needles can minimize this risk.
METHODS: The development of the suture delivery device for remote cardiac surgery requires: robust, precise and reliable suture placement; no manually loaded or loose needles; suture placement consistent with conventional wound closure techniques; no exposure of suturing needles except when intentionally deployed through the targeted tissue and excellent tissue visualization throughout device use. A customized device was designed, refined and tested to run 3-0 polypropylene suture for aortotomy closure in a variety of surgical tissue models in accordance with IACUC guidance.
RESULTS: The device proved reliable and ergonomic for closing aortotomy sites in more than 50 ex-vivo porcine aortas, the fully heparinized non-survivor bovine thoracic aorta model and in a survivor ovine model, with excellent intraoperative hemostasis under conditions of tachycardia, hypertension and full anticoagulation; histology at one week confirmed closure success. Use of the right anterior mini-thoracotomy cadaver model illustrated applicability to this minimal access approach. Throughout this testing, the device consistently retained the suturing needles in the unexposed, sheathed position until advanced by the surgeon. Since the integrated needles remain inside of the device, they do not need to be counted or placed in special post-use containers.
CONCLUSIONS: The study’s automated suture running device demonstrated excellent efficacy in a variety of remote aortotomy closure surgical models. The risk of inadvertent needle-stick injury appeared substantially reduced. This ergonomic and effective device encourages further evaluation towards a safer means of suturing in cardiac surgery, especially for minimal access aortotomy closure.