ISMICS 15 ISMICS 15 ISMICS 15
Exhibitors & Sponsors
 
 
Past Meetings
Future Meetings

Back to 2015 Annual Meeting Posters


Robot-Assisted Real-time Magnetic Resonance Image Guided Transcatheter Aortic Valve Replacement
Justin G. Miller, MD1, Ming Li, PhD1, Dumitru Mazilu, PhD1, Timothy Hunt, BS2, Keith A. Horvath, MD1.
1CSRP/NHLBI/NIH, Bethesda, MD, USA, 2LAMS/NIH, Bethesda, MD, USA.

OBJECTIVE: Real-time magnetic resonance image (rtMRI) guidance is proposed to be the future of transcatheter aortic valve replacement (TAVR) due to improved visualization, real time imaging, and pinpoint accuracy with device delivery. Unfortunately, performing a TAVR in a MRI scanner can be a difficult task due to limited space and a potentially awkward working environment. Our solution was to design a MRI compatible robot-assisted device that was capable of inserting and deploying a self-expanding valve from a remote computer console. We present our preliminary results using this device to perform transapical TAVR in a swine model.
METHODS: Using a commercially available 5 degrees of freedom (DoF) MRI compatible robotic arm, a 3 DoF valve delivery module was designed and developed. Via a subxiphoid incision and through apical purse strings, a 12mm trocar was inserted and secured in place. The delivery device and nitinol stented prosthesis were inserted into the trocar and mounted on the robotic delivery module. Two continuous real time imaging planes provided a virtual real time 3D reconstruction. The valve was remotely deployed by the surgeon via a graphic user interface.
RESULTS: In an acute, non-survival study, swine underwent robot-assisted rtMRI TAVR for evaluation of feasibility. Under rtMRI guidance the device deployment took approximately 60 seconds. Post-deployment necropsy was performed to confirm correlation between imaging and the actual valve position.
CONCLUSIONS: These results demonstrate the feasibility of robotic assisted TAVR using rtMRI guidance. This approach may eliminate some of the challenges of performing a procedure while working inside of an MRI scanner and may improve the success of TAVR. It provides superior visualization during the insertion process, deployment with pinpoint accuracy, and potentially communication between the imaging device and the robotic module which will prevent incorrect or misaligned deployment.


Back to 2015 Annual Meeting Posters