Prediction Of Perpendicular Projection For Transcatheter Aortic Valve Implantation Using A New Automatic 3D Computed Tomography Analyzing-Tool
Arnaud Van Linden, Johannes Blumenstein, Helge Möllmann, Won-Keun Kim, Tibor Ziegelhöffer, Christian Hamm, Thomas Walther, Jörg Kempfert.
Kerckhoff Klinik, Bad Nauheim, Germany.

OBJECTIVE: Different imaging techniques, including preoperative multi slice computed tomography (MSCT) images and aortic root angiography, have been suggested for predicting the optimal perpendicular C-arm angulation for transcatheter aortic valve implantation (T-AVI). The prototype software Aortic ValveGuide (AVG) of the DynaCT workstation (Siemens Healthcare, Germany) also allows for automatic processing of conventional multi slice computed tomography (MSCT) images. Aim of this study was to analyze predicted C-arm angulation using AVG compared to our institutional gold-standard 3mensio Valves™ (3mensio Medical Imaging BV, Netherlands) in preoperative MSCT.
METHODS: 158 preoperative MSCT image-datasets of patients undergoing T-AVI were included (66% transapical). MSCT raw-datasets were analyzed using AVG (Figure) software prototype with automatic detection of the relevant landmarks of the aortic root and using our gold-standard 3mensio Valves™. Two different perpendicular angulations were determined and compared. One angulation showed the left main on the right side of the screen (AngleA) and the other angulation had the left/right commissure in the middle (AngleB).
RESULTS: Automatic landmark detection worked well in all image-datasets. Minor manual adjustments were necessary in 29% of the patients. Mean AngleA for AVG was LAO/CRA 31±16/17±13° and for 3mensio 31±16/16±10° (p=0.3). Mean AngleB for AVG was LAO/CRA 43±11/30±10° and for 3mensio 43±11/38±7° (p<0.05. Mean deviation independent from CRA/CAU direction was 4° with a significant correlation between the two methods (r=0.7, respectively, p<0.001).
CONCLUSIONS: Aortic ValveGuide assessment of preoperative MSCT-scans seems to equally predict perpendicular C-arm angulation for T-AVI compared to 3mensio Valves™. Automatic processing of the image-datasets reduces the effort for MSCT analysis and may lead to more reproducible prediction. The accuracy seems to be patient-dependent rather than method-dependent.