OBJECTIVE: Transcatheter aortic valve implantation (TAVI) has shown promising results in patients with aortic valve disease at high risk for conventional heart surgery. The purpose of this study is to evaluate the safety and efficacy of transapical aortic valve implantation using the J-Valve™- a second-generation TAVI device.
METHODS: 15 patients with high risk aortic valve disease were enrolled in this study including 8 severe aortic stenosis (AS) and 7 severe pure aortic regurgitation (AR) (Mean age 74.7±4.3 years and Log-Euroscore 24.5±7.3%). Transapical TAVI procedure was done for these patients using the J-ValveTM - The Key features of the device are three “U-shape grasper” facilitating intuitive ‘self-positioning’ during valve implantation. Balloon pre-dilation was done as for patient with severe aortic stenosis. Clinical, echocardiographic as well CT angiogram evaluations were performed at baseline, post-procedure, discharge and 30 days.
RESULTS: TAVI with the J-Valve™ device was successful in 14 patients (procedural success rate 93%) with mean aortic valve annular size 23.8±1.7mm and prosthesis size 25.3±1.4mm. One patient received single stage PCI+TAVI due to severe aortic stenosis complicated with 80% stenosis of left coronary artery. No mortality was noted during the study period. Conversion to surgery was necessary in 1 patients with AS due to valve dislocation into the ascending aorta. No coronary obstructions were seen in study cohort. Post-procedure TAVI resulted in favorable reduction of mean transvalvular gradients as for patient with AS (57.2±7.9 vs. 17.8±10.7mmHg, P < 0.01). Mean transvalvular gradients as for patient with pure AR was 6.3±2.7mmHg. Paravalvular regurgitation immediately after implantation was graded none (n = 6), trace (n =8) which remains the same during 30 days follow-up. Complete left bundle branch block was noted in 4 patients after surgery. No other major complication was noted.
CONCLUSIONS: Trans-apical TAVI using J-Valve™ has the potential to become a safe and effective treatment option for both severe AS and AR at high risk for surgery.