Would any mitral annuloplasty ring and band be suitable for valve-in-ring technique ?
Gry Dahle, Kjell Arne Rein.
Rikshospitalet, OUS, Oslo, Norway.
OBJECTIVE: Mitral annuloplasty rings may be complete or incomplete and rigid, semi-rigid or flexible. Mitral valve repairs may fail and valve-in-ring (VIR) may be an option. So far there are a few reports for the Melody valve (Medtronic,Minneapolis, MN) and more frequent the Edwards Sapien/Edwards Sapien XT ((Edwards lifescinces, Irvine, CA) . The question is wether the valve will fit into and how it will function in a failed ring repair
METHODS: A bench test was done with selected annuloplasty rings with the Edwards Sapien XT deployed. Video and photos documented the procedure. The CE Edwards Classic rigid annuloplasty ring (Edwards lifescinces, Irvine, CA), Edwards semi-rigid Physioring (Edwards lifescinces, Irvine, CA), CG future band/ring (Medtronic, Minneapolis, MN) and the Profil 3D (Medtronic, Minneapolis, MN) were tested.
RESULTS: The CE Classic rigid ring was opened by the valve deployment and the Sapien XT was deformed, hence the leaflets were not sealing completely.
For the complete semi-rigid rings the rings turns out circular and the valve leaflets were sealing.
The incomplete semi-rigid CG future band was dislocated on the valve by deployment and not sealing. The complete semi-rigid CG future ring was not sealing complete the Edwards Sapien XT valve.
The rigid Profil 3D ring was deforming the Edwards Sapien XT and the leaflets were not sealing.
CONCLUSIONS: The incomplete rings will dislocate and may cause annulus rupture in real life. The rigid rings will deform the catheter valve, causing paravalvular leak and dysfunction of the valve leaflets. VIR is feasible for semi-rigid annuloplasty rings, though the right size must be selected to fit the valve to prevent dislocation or paravalvular leak.