A bubble-mixture method to improve dynamic image of carbon dioxide angiography
Te-I Chang, Sr., I-Hui Wu, Sr., Chih-Yang Chan, Sr..
National Taiwan University Hospital, Taipei, Taiwan.

OBJECTIVE: To demonstrate a novel way of preparing carbon dioxide (CO2) used for angiography by bubble-mixture of blood and CO2 gas before injection.
METHODS: CO2 angiography has provided a valuable option for patients with kidney diseases in endovascular procedures. Due to the buoyancy of CO2 gas, it preferentially fills nondependent vessels. The force of buoyancy may exceed the kinetic force of blood flow in vessels, especially in smaller distal vessels, causing transient gas trapping and poor dissemination to distal vessels. By premixture of blood and CO2 with a bubble-creating technique, it allows a single CO2 volume to be dispersed into numerous tiny bubbles. In this way, the vapor-lock phenomenon can be dramatically reduced and the distribution of CO2 inside target vessels is more even, thus creating a more contrast-like image during digital subtraction angiography.
The bubble-mixing technique was very simple and safe by just adding one three-way lock and a syringe into previous closed CO2 delivery system. The proteins in the blood act as natural foaming agents. The blood was withdrawn from the patient by the same delivery catheter and by acting piston motions of two parallel syringes, we create bubble mixtures of CO2 and blood. The risk of air contamination was minimized because no parts of closed system was disconnected during preparation.
RESULTS: A patient with left superficial femoral artery occlusive lesion was presented. We delivered pure CO2 and bubble-mixed CO2 during angiographies for comparison. The pure CO2 subtraction images showed strolling of CO2 back and forth and poor enhancement of distal femoral artery due to low distal runoff before angioplasty, comparing to that, the bubble-mixed CO2 images showed smooth progression of CO2 and better visualization of distal arteries. After balloon angioplasty, the distal run off improved.
CONCLUSIONS: The dynamic image quality change for CO2 digital subtraction angiography was promising. It solves uneven distribution of CO2 toward antigravity side of vessels and provides a more smooth “contrast-like” dynamic images of CO2 angiography.