A Novel Heart Valve Of Medical-grade Ultra High Molecular Weight Polyethylene Fibers; In Vitro Blood Cell Adhesion experiments
Amir Basir1, Mark Roest1, Wally Müller2, Joost van Herwaarden1, Frans Moll1, Gerard Pasterkamp1, Jolanda Kluin1, Paul Gründeman1, Philip de Groot1.
1University Medical Center Utrecht, Utrecht, Netherlands, 2Utrecht University, Utrecht, Netherlands.

OBJECTIVE: Different prostheses materials are available for vascular- and heart valve replacement. All of them carry more or less serious drawbacks; some are not suitable for endovascular treatment by their inability to fold or have too large profile, other need strong anticoagulant drugs or have a moderate durability. Medical-grade Ultra High Molecular Weight Polyethylene (UHMWPE) fibers are very thin, flexible, fatigue- and abrasion resistant and have high strength. Therefore, prostheses made of these fibers might be attractive for construction of heart valves applied in transcatheter approaches. The aim of this study is to test the blood cell adhesion on woven patch of UHMWPE and to compare this with the most common cardio-vascular patches.
METHODS: A UHMWPE patch and five commonly used patches of ePTFE and polyester were perfused with human blood of three healthy donors in a double flat rectangular perfusion chamber connected to a roller pump. The perfusion was performed six times per donor. During each perfusion session two different patches were perfused with 15 ml whole blood during three minutes. After the perfusion nine different pictures of selected areas were taken from the middle, the top and the bottom of each patch using scanning electron microscopy and were assessed for different types of cell adhesion.
RESULTS: Less platelet-aggregates were seen per field on UHMWPE patch compared to the woven polyester patch (P= 0.01). Compared to the other polyester and ePTFE patches there was not a significantly difference for UHMWPE patch regarding adherence of platelet-aggregation. For red blood cell adherence there was no significant differences for these patch compared to the five patches.
CONCLUSIONS: UHMWPE patch hold promise to be a potentially suitable material in blood for prosthetic valve and other low profile cardio-vascular applications by their non-inferiority regarding platelet aggregation compared to test materials. Yet, more extensive In Vivo studies are needed to confirm if medical-grade UHMWPE is a suitable candidate for use in prosthetic heart valves and other vascular applications.