Computational Tool to Measure the Length of Artificial Chordae for Key Hole Mitral Valve Repair
Abdullrazak Hossien.
Morriston University Hospital, Swansea, United Kingdom.
Objective:
Non-resection technique is considered a valuable procedure in treatment ofmitral valve (MV) incompetence using artificial chordae. The length of artificial chordae is a crucial step in the creation of optimal coaptation of leaflets. We propose a novel method to measure the height of chordae.
Methods:
We processed the 3D images obtained from Phillips 3D transesophageal Echocardiography(3D TOE) for normal and prolapse mitral valve. The images were cropped by Qlab between the origin of chords on the papillary muscle and their insertion on the mitral leaflets in the mid and late systole phase of cardiac cycle. The cropped model was sliced and segmented computationally by using open source software program and mathematical equations to build a 3D Chordae model.
Results:
A high quality image set of the chords was retrieved from 3D TOE. The slicing of images results in 7 images of 0.35 mm thickness. The reconstruction of the slices results in 3D anatomical segmentation of the Chords. Binary STL file (STereoLithography) was built by processing the segmented MV model which was meshed as well. 3D segmentation of the chords results in understanding the geometry of leaflet coaptation, distribution of chords and variation of chordal length in relation to individual position. the length measurements were 2,43 cm +- 3,5 mm.
Conclusion:
Computational measurement is a helpful and valuable tool in predicting chordae length for surgical planning of mitral valve repair and may play an important roll in key hole techniques mitral valve surgery.